Preparing for Use
R&S
®
SMBV100A
54
Operating Manual 1176.8016.02 ─ 17
For information about the LXI standard, refer to the LXI website at
. See also "News from Rohde & Schwarz, article 2006/II - 190".
The R&S
SMBV provides an integrated "LXI Status" dialog for LXI status indication and
reset of the LAN configuration ("LAN Configuration Initialize", LCI). To access the LXI
status dialog, press the SETUP key and select "Remote > LXI Status".
For further information, see section "LXI Status" in the Operating Manual.
Firmware update
After a firmware update, shut down and restart the instrument in order to enable the full
LXI functionality.
Default state of the network settings
According to the LXI standard, an LCI must set the following parameters to a default
state.
Parameter
Value
TCP/IP Mode
DHCP + Auto IP Address
Dynamic DNS
Enabled
ICMP Ping
Enabled
Password for LAN configuration
LxiWebIfc
The LCI for the R&S
SMBV also resets the following parameters:
Parameter
Value
Hostname
<Instrument-specific host name>
Description
Vector Signal Generator
Negotiation
Auto Detect
VXI-11 Discovery
Enabled
The LAN settings are configured using the instrument's "LXI Browser Interface".
2.8.1 LXI Browser Settings
To access the instrument via the web browser:
► Type in the instrument's host name or IP address in the address field of the
browser on your PC, for example "http://10.113.1.205".
LXI Configuration