Remote Control Commands
R&S
®
SMBV100A
658
Operating Manual 1176.8016.02 ─ 17
Return values:
<RMode>
STOP | RUN
RUN
The signal is generated. A trigger event occurred in the triggered
mode.
STOP
The signal is not generated. A trigger event did not occur in the
triggered modes, or signal generation was stopped by the com-
mand
:BB:DM:TRIG:ARM:EXECute
(armed trigger modes
only).
Example:
BB:DM:TRIG:SOUR EXT
sets external triggering
BB:DM:TRIG:MODE ARET
selects the Armed_Retrigger mode
BB:DM:TRIG:RMOD?
queries the current status of signal generation.
Response: RUN
the signal is generated, an external trigger was executed.
Usage:
Query only
Manual operation:
See
[:SOURce<hw>]:BB:DM:TRIGger:SLENgth
<SLength>
Defines the length of the signal sequence to be output in the "Single" trigger mode.
Parameters:
<SLength>
integer
Range:
1 to 4294967295
*RST:
1000
Default unit: symbol
Example:
BB:DM:SEQ SING
sets trigger mode Single.
BB:DM:TRIG:SLEN 200
sets a sequence length of 200 symbols. 200 symbols will be out-
put after the next trigger event.
Manual operation:
See
[:SOURce<hw>]:BB:DM:TRIGger:SOURce
<Source>
Selects the trigger source.
SOURce Subsystem