Remote Control Commands
R&S
®
SMBV100A
629
Operating Manual 1176.8016.02 ─ 17
Parameters:
<Step>
float
Range:
-80 to 80
Increment: 0.01
*RST:
0
Default unit: dB
Example:
BB:ARB:MCAR:EDIT:CARR:POW -80dB
sets a power of -80 dB for the carriers in the carrier range.
BB:ARB:MCAR:EDIT:CARR:POW:STEP 1 dB
the power is incremented by 1dB for each carrier, i.e. the first
carrier has -80dB, the second -79dB, etc.
Manual operation:
See
[:SOURce<hw>]:BB:ARBitrary:MCARrier:EDIT:CARRier:POWer[:STARt]
<Start>
The command sets the power for the individual carriers in the defined carrier range. If
the command
:BB:ARB:MCAR:EDIT:CARR:POW:STEP
is used to define a step width,
the power entered here applies only to the starting carrier. The power of the remaining
carriers is stepped up or down by the power specified in
the
:BB:ARB:MCAR:EDIT:CARR:POW:STEP
command.
Parameters:
<Start>
float
Range:
-80 to 0
Increment: 0.01
*RST:
0
Default unit: dB
Example:
BB:ARB:MCAR:EDIT:CARR:POW -50 dB
sets the power of the carriers in the carrier range to -50 dB.
Manual operation:
See
[:SOURce<hw>]:BB:ARBitrary:MCARrier:EDIT:CARRier:STARt
<Start>
[:SOURce<hw>]:BB:ARBitrary:MCARrier:EDIT:CARRier:STOP
<Stop>
The command selects the last carrier in the carrier range to which the settings with
the
:BB:ARB:MCAR:EDIT:CARR:..
commands shall apply.
Parameters:
<Stop>
integer
Range:
0 to 511
*RST:
0
Example:
BB:ARB:MCAR:EDIT:CARR:STOP 4
the carrier range stops at carrier 4.
Manual operation:
See
SOURce Subsystem