![Rohde & Schwarz SMBV100A Operating Manual Download Page 610](http://html1.mh-extra.com/html/rohde-and-schwarz/smbv100a/smbv100a_operating-manual_1477735610.webp)
Remote Control Commands
R&S
®
SMBV100A
595
Operating Manual 1176.8016.02 ─ 17
[:SOURce<hw>]:AWGN:FREQuency:RESult?
In CW inerferer mode, queries the actual frequency of the sine.
The actual frequency may differ from the desired frequency, since the resolution is limi-
ted to 0.7 Hz.
Return values:
<Result>
float
Range:
-40E6 to 40E6
Increment: 0.01
*RST:
0
Example:
AWGN:FREQ:RES?
queries the actual frequency of the interfering sine
Usage:
Query only
Manual operation:
See
" Resulting CW Frequency - AWGN "
[:SOURce<hw>]:AWGN:FREQuency:TARGet
<Target>
Sets the desired frequency of the sine in
AWGN:MODE CW
mode.
The resulting frequency may differ from the desired frequency because of the limited
frequency resolution of 0.7 Hz.
Parameters:
<Target>
float
Range:
-40E6 to 40E6
Increment: 0.01
*RST:
0
Example:
AWGN:FREQ:TARG 2kHz
sets a frequency of 2 kHz for the interfering sine.
Manual operation:
See
" Target CW Frequency - AWGN "
[:SOURce<hw>]:AWGN:MODE
<Mode>
Selects the mode for generating the interfering signal.
Parameters:
<Mode>
ONLY | ADD | CW
ADD
The AWGN noise signal is added to the baseband signal.
ONLY
The pure AWGN noise signal is modulated to the carrier. The
connection to the baseband is interrupted.
CW
The sine interfering signal is added to the baseband signal.
*RST:
ADD
SOURce Subsystem