Remote Control Commands
R&S
®
SMBV100A
578
Operating Manual 1176.8016.02 ─ 17
:OUTPut<hw>[:STATe]
<State>
Activates and deactivates the RF output signal (RF ON / RF OFF).
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
0
Example:
OUTP OFF
deactivates the RF output.
Manual operation:
See
:OUTPut<hw>[:STATe]:PON
<Pon>
Selects the state of the RF output when the instrument is switched on.
Parameters:
<Pon>
OFF | UNCHanged
OFF
deactivates the output when the instrument is switched on (RF
OFF).
UNCHanged
restores the initial state of the RF output before the last turn off.
sets the output status as it was when the instrument was
switched off.
*RST:
n.a. (factory preset: UNCHanged)
Example:
OUTP:PON OFF
RF output A is deactivated when the instrument is switched on.
Manual operation:
See
" Power-On State - RF Signal "
7.15 SENSe, READ and INITiate Subsystems
The SENSe subsystem contains the commands for configuring the power measure-
ments with R&S NRP-Zxx power sensor(s) connected to the generator. The measure-
ment is started and the measurement result retrieved with the
READ
command. The
description of this commands is included in the following.
Up to four sensors can be connected to the signal generator. They are distinguished by
means of the suffix under SENSe:
●
Power sensor connected to the SENSOR port =
SENSe[1]
●
Power sensor connected first to one of the USB interfaces =
SENSe 2
●
Further power sensors connected second and third via USB =
SENSe 3
and
SENSe 4
The suffixes <ch>
SENSe, READ and INITiate Subsystems