Remote Control Commands
R&S
®
SMBV100A
541
Operating Manual 1176.8016.02 ─ 17
Parameters:
<Merror>
integer
Range:
0 to 4294967295
*RST:
100
Example:
BERT:SET:MERR 1E4
sets a measurement over 1000 errors.
Manual operation:
See
:BERT:SETup:RESTart[:STATe]
<State>
Activates/deactivates an external restart of the BER measurement.
Parameters:
<State>
INTernal | EXTernal
*RST:
INTernal
Example:
BERT:SET:REST:STAT INTernal
The external signal restarts the measurement.
Manual operation:
See
:BERT:SETup:TYPE
<Type>
Selects the PRBS sequence. The data generated by the PRBS generator is used as a
reference for the measurement.
Parameters:
<Type>
PRBS9 | PRBS11 | PRBS15 | PRBS16 | PRBS20 | PRBS21 |
PRBS23
*RST:
PRBS9
Example:
BERT:SET:TYPE PN15
sets a pseudo random binary sequence consisting of 32767 bits.
Manual operation:
See
:BLER:SETup:TYPE
<Type>
Queries the CRC polynomial used. CCITT CRC 16 : G(x) = x
16
+ x
12
+ x
5
+ x
1
is the
CRC polynomial supported.
Parameters:
<Type>
CRC16
*RST:
CRC16
Example:
BLER:SET:TYPE
queries the type of measurement.
Example:
Response: CRC16
Manual operation:
See
BERT and BLER Subsystems