![Rohde & Schwarz SMBV100A Operating Manual Download Page 534](http://html1.mh-extra.com/html/rohde-and-schwarz/smbv100a/smbv100a_operating-manual_1477735534.webp)
Remote Control Commands
R&S
®
SMBV100A
519
Operating Manual 1176.8016.02 ─ 17
Example:
{MARKER LIST 1: 0:0;10:1;20:0;30:1}
BB:DM:CLIS:TAG? 'MARKER LIST 1'
Queries the content of the
MARKER LIST 1
tag of the selected
control list file.
Response: '0:0;10:1;20:0;30:1'
The marker setting for samples 0 to 9 = 0 (low), for 10 to 19 = 1
(high) and for 20 to 29 = 0. From sample 30 onward the marker
setting = 1.
Example:
{LEVATT LIST 1: 0:0;10:1;20:0;30:1}
BB:DM:CLIS:TAG? 'LEVATT LIST 1'
Queries the content of the
LEVATT LIST 1
tag of the selected
control list file.
Response: '0:0;10:1;20:0;30:1'
Level attenuation applies to data values 10 to 19 (high) and from
data value 30 onward.
Usage:
Setting only
{WAVEFORM-Length: #I0Q0I1Q1...IxQx...IN-1QN-1…}
(mandatory for waveforms)
The tag contains the actual waveform data or multi-segment waveform data (I/Q
stream).
Use the
tag to define whether the waveform file contains a normal waveform or
a multi-segment waveform.
The following figure shows the principle of multi-segment work mode.
Waveform and List Format