Instrument Function
R&S
®
SMBV100A
445
Operating Manual 1176.8016.02 ─ 17
"Retrigger"
The signal is generated continuously. A trigger event (internal or
external) causes a restart.
"Armed_Auto"
The signal is generated only when a trigger event occurs. Then the
signal is generated continuously.
Button "Arm" stops signal generation. A subsequent trigger event
(internal with "Execute Trigger" or external) causes a restart.
"Armed_Ret-
rigger"
The signal is generated only when a trigger event occurs. Then the
signal is generated continuously. Every subsequent trigger event cau-
ses a restart.
Button "Arm" stops signal generation. A subsequent trigger event
(internal with "Execute Trigger" or external) causes a restart.
"Single"
The signal is generated only when a trigger event occurs. Then the
signal is generated once to the length specified at "Signal Duration".
Every subsequent trigger event (internal with "Execute Trigger" or
external) causes a restart.
Remote command:
[:SOURce<hw>]:BB:MCCW[:TRIGger]:SEQuence
Signal Duration
Defines the length of the signal sequence to be output in the "Single" trigger mode.
The unit of the entry is defined under "Signal Duration Unit". It is then possible to out-
put deliberately just part of the signal, an exact sequence of the signal, or a defined
number of repetitions of the signal.
Remote command:
[:SOURce<hw>]:BB:MCCW:TRIGger:SLENgth
Running/Stopped
Displays the status of signal generation for all trigger modes. This display appears only
when signal generation is enabled ("State" On).
"Running"
The modulation signal is generated; a trigger was (internally or exter-
nally) initiated in triggered mode.
If "Armed_Auto" and "Armed_Retrigger" have been selected, genera-
tion of signals can be stopped with the "Arm" button. A new trigger
(internally with "Execute Trigger" or externally) causes a restart.
"Stopped"
The signal is not generated, and the instrument waits for a trigger
event (internal or external).
Remote command:
[:SOURce<hw>]:BB:MCCW:TRIGger:RMODe?
Arm
Stops signal generation. This button appears only with "Running" signal generation in
the "Armed_Auto" and "Armed_Retrigger" trigger modes.
Signal generation can be restarted by a new trigger (internally with "Execute Trigger"
or externally).
Remote command:
[:SOURce<hw>]:BB:MCCW:TRIGger:ARM:EXECute
Baseband Signal - Baseband Block