![Rohde & Schwarz SMBV100A Operating Manual Download Page 372](http://html1.mh-extra.com/html/rohde-and-schwarz/smbv100a/smbv100a_operating-manual_1477735372.webp)
Instrument Function
R&S
®
SMBV100A
357
Operating Manual 1176.8016.02 ─ 17
The menu offers only the coding settings that are permissible for the chosen modula-
tion type. The other coding methods are grayed out.
If the system is subsequently switched to a modulation type for which the selected cod-
ing is not available, coding is automatically set to OFF.
Remote command:
Modulation
The modulation type is set in the "Modulation section". The parameters offered depend
on what is currently selected. The "More..." button opens a submenu for displaying the
user-defined mapping.
Modulation Type
Selects a modulation type.
The associated symbol mapping is displayed in the "More..." submenu (see
ter 5.9.3.2, "Modulation Types - Custom Digital Mod"
If the "coding" that is set is not possible with the chosen modulation type, "coding" is
set to "Off".
For selection "Variable FSK", the deviation of each symbol can be set in the "More..."
submenu.
Remote command:
Load User Mapping
Selects user defined mapping table. This opens the "Select List File User Mapping"
window in which the mapping table can be selected (see
on page 342). The button is only available if modulation type "USER" is selected.
The button "File Manager" calls the "File Manager" menu. The "File" "Manager" menu
is used to display and delete files.
Remote command:
[:SOURce<hw>]:BB:DM:MLISt:SELect
[:SOURce<hw>]:BB:DM:MLISt:CATalog?
[:SOURce<hw>]:BB:DM:MLISt:DELete
ASK Depth
Sets the modulation depth m for ASK modulation.
m = (Amplitude
max
-Amplitude
min
) / (Amplitude
max
+Amplitude
min
)
Remote command:
FSK Deviation
Sets the frequency deviation for FSK modulation. The range of values depends on the
chosen symbol rate (see data sheet).
Whenever "MSK" is selected, the deviation corresponds to 1/4 of the symbol rate and
cannot be set.
Baseband Signal - Baseband Block