Impairments
I/Q impairment for specific impairment of the I/Q modulation is set in the "Impairments"
section.
State - Impairments I/Q Mod
Activates/deactivates I/Q impairment.
If activated, the settings for leakage, I/Q imbalance and quadrature offset become
effective.
Whether or not I/Q impairment is activated is indicated in the function block.
Note:
It is advisable to switch automatic level control to the Sample&Hold mode, as
otherwise level errors may occur (see
Chapter 5.5.7, "Automatic Level Control - ALC"
Remote command:
[:SOURce<hw>]:IQ:IMPairment[:STATe]
Offset - I/Q Mod
Sets the carrier offset (in percent) of the amplitudes (scaled to the peak envelope
power (PEP) for the I and/or Q signal component. An ideal I/Q modulator suppresses
the carrier offset completely (offset = 0 percent).
For more information, see
Chapter 5.7.1.2, "I and Q Offset"
Remote command:
[:SOURce<hw>]:IQ:IMPairment:LEAKage:I
[:SOURce<hw>]:IQ:IMPairment:LEAKage:Q
Gain Imbalance - I/Q Mod
Sets the imbalance of the I and Q vector (see
Chapter 5.7.1.1, "Gain and Gain Imbal-
The entry is made in dB (default) or %, where 1 dB offset is roughly 12 % according to
the following:
Imbalance [dB] = 20log ( | GainQ | / | GainI |)
Positive values mean that the Q vector is amplified more than the I vector by the corre-
sponding percentage. Negative values have the opposite effect.
Remote command:
[:SOURce<hw>]:IQ:IMPairment:IQRatio[:MAGNitude]
Quadrature Offset - I/Q Mod
Sets the quadrature offset (see
Chapter 5.7.1.3, "Quadrature Offset"
Remote command:
[:SOURce<hw>]:IQ:IMPairment:QUADrature[:ANGLe]
5.6.2.3
I/Q Swap and I/Q Wideband
The I/Q control can be swapped and the optimized setting for wideband signals can be
selected in the middle section of the menu.