Instrument Function
R&S
®
SMBV100A
291
Operating Manual 1176.8016.02 ─ 17
Remote command:
Dwell Time - List Mode
Enters the dwell time. The dwell time determines the duration of a list step in list oper-
ating modes "Auto", "Single" and "Extern Single". In these modes a complete list is pro-
cessed either once or continuously.
In list operating modes "Step" and "Extern Step", the set dwell time does not affect sig-
nal generation. In this case, the duration of a list step is determined by the time
between two (internal or external) trigger events.
The "Dwell Time" set by the user is used as the step time of the list mode. The effec-
tive net dwell time is shorter, reduced by the setting time. This setting time may be
greater than the time specified in the data sheet.
Remote command:
Current Index - List Mode
Sets the list index in "Step" mode.
Remote command:
Learn List Mode Data... - List Mode
Starts the determination of the hardware setting for the selected list. The data deter-
mined in this way is stored along with the list.
It may be necessary to deliberately activate list learning in the event of greatly altered
environmental conditions that require new hardware settings.
If this is not done, a previously learned hardware setting will continue to be used when
list mode is switched on ("State = On"). If no setting is available, e.g. when the list is
used for the first time, learning is automatically activated.
Remote command:
RF Block