![Rohde & Schwarz SMBV100A Operating Manual Download Page 245](http://html1.mh-extra.com/html/rohde-and-schwarz/smbv100a/smbv100a_operating-manual_1477735245.webp)
Instrument Function
R&S
®
SMBV100A
230
Operating Manual 1176.8016.02 ─ 17
5.5.8.1
NRP-Z Level Control Settings
► To access the dialog for configuring the level control settings, perform one of the
following:
● Select "RF/A Mod > config... > RF Level > NRP-Z Level Control".
● Press the MENU key and select "RF/A Mod > RF Level > NRP-Z Level Con-
trol".
The dialog contains all parameters for configuring the settings for level control test
setup.
The remote commands required to define these settings are described in
ter 7.16.17, "SOURce:POWer Subsystem"
Sensor
Selects the R&S NRP-Z power sensor for power control.
Note:
In remote control, the sensors are set up using the
SENSe
commands. The
remote measurement is triggered by the
READ
query which also provides the measure-
ment results.
The software version of the connected power sensor can be retrieved by means of the
remote control command
:SENS:POW:TYPE?
.
Use the "Setup >"
Chapter 5.2.3.5, "NRP-Z Info/Update"
the sensor software.
Remote command:
[:SOURce<hw>]:POWer:SPC:SELect
State
Activates power control using the selected sensor.
The control loop periodically adjusts the generator output. After switching off, the run-
ning loop is completed.
Remote command:
Measured Level
Indicates the current reading of the sensor.
RF Block