The level entered with this command corresponds to the level at the RF output, i.e. any
offset entry is not taken into consideration.
Note:
The SCPI command
sets the level of the "Level" display, i.e. the level containing offset.
Parameters:
<Power>
Minimum level ... Maximum level
The value range for the level setting varies according to the instru-
ment model.
The values are given in the data sheet.
Increment: 0.01 dB
*RST:
-30 dBm
Default unit: dBm
Example:
SOUR:POW:POW 15
sets the RF level at output to 15 dBm.
[:SOURce<hw>]:POWer:SPC:CRANge
<PowCntrlCRange>
Defines the capture range of the power control system.
Within the range:
Target Level +/- Catch Range
the power control locks and tries to achieve the target level. Readings outside the range
are not considered.
Parameters:
<PowCntrlCRange> float
Range:
0 to 50
Increment: 0.01
*RST:
30
Default unit: dB
Example:
POW:SPC:CRAN 15
sets the capture range to +/- 15 dB.
[:SOURce<hw>]:POWer:SPC:DELay
<PowCntrlDelay>
Defines a waiting period between the level adjustment of the generator and the next
measurement of the power sensor.
Parameters:
<PowCntrlDelay>
integer
Range:
0 to 1000
Increment: 1
*RST:
0
Example:
:POW:SPC:DEL 2 ms
the sensor starts the next reading 2 ms after the level adjustment.