Instrument Function
R&S
®
SMBV100A
375
Operating Manual 1407.6062.32 ─ 08
The carrier power that is set with
Power + n* Power Step
must be within the valid value
range -80 dB to 0 dB.
SCPI command:
Phase Start
Sets the phase of the starting carrier. This setting is only available for "Optimize Crest
Factor Mode" = Off.
SCPI command:
Phase Step
Sets the width of the step with which the phase will be changed from carrier to carrier.
The phase that is set with
Phase + n* Phase Step
must be within the valid value range
0° to 360°.
This setting is only available for "Optimize Crest Factor Mode" = Off.
SCPI command:
Accept
Adopts the carrier range setting into the table ("Carrier Table").
SCPI command:
Carrier Table and Carrier Graph
The lower part of the menu is used to open submenus for configuring carrier tables and
for setting triggers, markers and clocks.
Carrier Table
Calls the table for configuring individual carriers. This configuration can be checked with
the aid of the "Carrier Graph".
The table displays the settings of all available carriers. Carriers in the On state are high-
lighted. All carrier parameters can be edited in the table.
Baseband Signal - Baseband Block