Instrument Function
R&S
®
SMBV100A
230
Operating Manual 1407.6062.32 ─ 08
The "Dwell Time" set by the user is used as the step time of the list mode. The effective
net dwell time is shorter, reduced by the setting time. This setting time may be greater
than the time specified in the data sheet.
SCPI command:
Current Index - List Mode
Sets the list index in "Step" mode.
SCPI command:
Learn List Mode Data... - List Mode
Starts the determination of the hardware setting for the selected list. The data determined
in this way is stored along with the list.
It may be necessary to deliberately activate list learning in the event of greatly altered
environmental conditions that require new hardware settings.
If this is not done, a previously learned hardware setting will continue to be used when
list mode is switched on ("State = On"). If no setting is available, e.g. when the list is used
for the first time, learning is automatically activated.
SCPI command:
List Mode Data... - List Mode
Calls the "File Select" menu for selecting and creating a list or the "File Manager".
SCPI command:
RF Signal and Analog Modulations - RF Block