"Step"
Manual, step-by-step processing of the list. Activating "Step" stops the
current list and the cursor moves to the value displayed for "Current
Index". It is now possible to scroll up and down in the list in discrete
steps by varying the index. The duration of a list step is determined by
the time between two index entries.
Button "Reset" restarts the list at the starting point.
"Extern Single" Single cycle from the beginning to the end of the list as with "Single",
but started by an external trigger.
The external trigger signal is input at the BNC connector INST TRIG.
Button "Reset" restarts the list at the starting point.
"Extern Step"
Step-by-step cycle using the external trigger signal. Each trigger event
starts a single step. The duration of a list step is determined by the time
between two trigger events.
The external trigger signal is input at the BNC connector INST TRIG.
Button "Reset" restarts the list at the starting point.
SCPI command:
Execute Single - List Mode
Triggers the list manually. This button is available only if mode "Single" is selected.
SCPI command:
Reset - List Mode
Resets the list to the starting point.
SCPI command:
Dwell Time - List Mode
Enters the dwell time. The dwell time determines the duration of a list step in list operating
modes "Auto", "Single" and "Extern Single". In these modes a complete list is processed
either once or continuously.
In list operating modes "Step" and "Extern Step", the set dwell time does not affect signal
generation. In this case, the duration of a list step is determined by the time between two
(internal or external) trigger events.