Instrument Function
R&S
®
SMBV100A
158
Operating Manual 1407.6062.32 ─ 08
Offset - RF Signal
Sets the frequency offset relative to the RF frequency. The frequency offset of a down-
stream instrument (e.g. a mixer) is entered.
The entry does not change the value of the RF frequency at the RF output. It only changes
the RF frequency displayed in the display header. The value of the RF frequency in the
header corresponds to the frequency at the output of the downstream instrument.
SCPI command:
Multiplier - RF Signal
Sets the multiplication factor for the RF frequency.
In the frequency field of the status bar, the instrument adjusts its frequency display
according to the set multiplication factor. This frequency value shows the frequency at
the output of the downstream multiplier. The entry does not change the RF frequency at
the RF output of the R&S SMBV.
SCPI command:
User Variation
If the frequency is set using the rotary knob, the step width is defined in the "User Var-
iation" section.
Variation Active - RF Signal
Activates the user-defined step width used when varying the frequency value with the
rotary knob.
"ON"
The frequency value set with the rotary knob is varied using the user-
defined step width which is entered under "Variation Step".
"OFF"
The frequency value set with the rotary knob is varied in steps of one
unit at the cursor position (standard operating mode).
SCPI command:
Variation Step - RF Signal
Sets the user-defined step width. This step width is used when entering the RF frequency
using the rotary knob. Frequency variation with this step width must also be activated
with "Variation Active".
SCPI command:
5.4.3 Phase
The phase of the RF output signal can be changed in the "Phase Settings" section of the
"RF Frequency/Phase" dialog.
RF Signal and Analog Modulations - RF Block