Firmware Update 

3.1  Validity Information 

Device 

Order Number 

R&S® SMB100B RF Analog Signal Generator

 

1422.7950K02

 

3.2  Downgrade 

Generally it is not recommended to use an earlier version than the latest version 
available. In some cases the older versions do not support the hardware used in your 
instrument. Before installing this firmware, check if this could happen:  

►  Start System Config

Setup

Instrument Assembly

Versions / Options 

In the tab “Firmware”, you find the  

 

►  If the version to be installed is greater or equal than the “Min.Version”, the 

hardware will be supported after downgrading. (However, this cannot be 
guaranteed for all software options) 

►  If the version to be installed is lesser than the “Min.Version”, not all of the modules 

will be supported. 

You instrument will not work after downgrading!

 

►  Downgrading may fail using standard rsu-Files (eg. due to changes in the 

instrument configuration file). In this case, press PRESET-Button during power-on 
or install ISO image available from service department. 

3.3  Update Information 

The update procedure requires that the instrument is operational. There is no need to 
uninstall the current firmware. Instrument settings are preserved during the update, 
including user data and network settings. 

 

 

 

 

 

To perform this procedure, USB Storage must be enabled in security settings. 
Press the 

SETUP

 key, select 

Security

 and check 

USB Storage

 setting. 

 

   

 

Summary of Contents for SMB100B

Page 1: ...com Subject to change R S is a registered trademark of Rohde Schwarz GmbH Co KG Trade names are trademarks of the owners 1422 7950 02 Version 05 R S SMB100B The software makes use of several valuable open source software packages For information see the Open Source Acknowledgment provided with the product The following abbreviations are used throughout this document R S SMB100B is abbreviated as R...

Page 2: ... 4 Version 4 50 074 45 5 1 5 Version 4 30 060 18 7 2 Modifications to the Documentation 8 3 Firmware Update 9 3 1 Validity Information 9 3 2 Downgrade 9 3 3 Update Information 9 3 4 Updating the Firmware 10 3 5 Alternative update procedures 11 3 5 1 Firmware update over LAN 11 3 5 2 Firmware update using ISO image 11 4 Customer Support 14 ...

Page 3: ...ral Secure Update Policy These settings are password protected For more information see the R S SMB100B User Manual 1 1 Version 4 70 128 48 New Options None New Functionality Changed Behavior Common Ca 30 faster internal adjustments 662773 Temperature Compensation for increased Level accuracy for temperatures below 18 C and above 33 C is active with datasheet Version higher than 6 0 365716 Fixed I...

Page 4: ...sed by pressing the rotary knob for some time 362578 Fixed Issues Common Hardcopy By default screen shots are not saved to USB memory 550938 Mouse operation disturbed after touching screen remote control via VNC does not work as expected 489231 RF In rare cases the error message RFB6 BB Output unleveled Gain Control on Lower Limit appears if pulse modulation is active 585193 AM Modulation The modu...

Page 5: ...6033 LF Output A new value Load Impedance in certain emulations 542234 Fixed Issues Common General Slide switches can be operated in remote state 501226 Internal Adjustment Improved Table Adjustment Internal Adjustments have to be done 562367 Remote Control Error handling SYSTEM SERROR delivers info messages different from SYSTEM ERROR 555983 Save Recall Frequency is recalled even in case of Exclu...

Page 6: ...or frequencies below 1 MHz 465312 List Mode The minimal dwell time has been reduced to 1 ms 455790 Sweep The minimal dwell time has been reduced to 1 ms 455790 Fixed Issues Common Remote Control After viClear remote connection via USB not working 380954 UCOR After executing Fill UCOR data with sensor error messages occures twice 422497 The correction value doesn t have an effect 458407 Known Issue...

Page 7: ...ce Input Output R S SMBB B5 Stereo RDS Coder R S SMBB B32 Ultra High Output Power R S SMBB B86 Remote Control Interface GPIB and USB R S SMBB K22 Pulse Modulator R S SMBB K23 Pulse Generator R S SMBB K24 Multifunction Generator R S SMBB K27 Pulse Train R S SMBB K704 Flexible Reference Input from 1 MHz to 100 MHz R S SMBB K720 AM FM PhiM New Functionality Changed Behavior Initial firmware version f...

Page 8: ...2 Modifications to the Documentation The current documentation is up to date ...

Page 9: ...owever this cannot be guaranteed for all software options If the version to be installed is lesser than the Min Version not all of the modules will be supported You instrument will not work after downgrading Downgrading may fail using standard rsu Files eg due to changes in the instrument configuration file In this case press PRESET Button during power on or install ISO image available from servic...

Page 10: ...on instrument if instrument is powered off Wait a few seconds until Process Software Update message box appears Confirm by touching the YES Button or pressing the rotary knob Select firmware version using the arrow keys and press knob to start update Wait until Software update successful message box appears This may take several minutes Remove USB stick and touch the Reboot button The instrument n...

Page 11: ... procedure starts immediately Execute internal adjustments if indicated 3 5 2 Firmware update using ISO image Potential loss of data User data and user specific instrument settings will be lost during this procedure Instrument serial number software license keys and all adjustments requiring external measuring equipment are not affected Required equipment Software ISO image for firmware update SMB...

Page 12: ...lect medium CD Select Create Data CD From the Files menu open file SMBB_ version number iso Click Burn When finished close Nero and disconnect external USB CD DVD drive Install new firmware on R S SMB100B Instrument must be switched off Connect the external USB CD DVD drive to the R S SMB100B Switch on Instrument The instrument boots from external drive Follow the instructions on screen Disconnect...

Page 13: ...ootin menu The ISO is now copied to the SMBV100B Start the installation If the CD refuses to boot please ensure that you have burned the ISO image as an image and not as a single file Check the CD regarding presence of several files ...

Page 14: ... to date and to be informed about new application notes related to your instrument please send an e mail to the Customer Support Center stating your instrument and your wish We will take care that you will get the right information Europe Africa Middle East Phone 49 89 4129 12345 customersupport rohde schwarz com North America Phone 1 888 TEST RSA 1 888 837 8772 customer support rsa rohde schwarz ...

Reviews: