![Rohde & Schwarz Smart Instruments UP300 Operating Manual Download Page 271](http://html1.mh-extra.com/html/rohde-and-schwarz/smart-instruments-up300/smart-instruments-up300_operating-manual_1477754271.webp)
R&S UP300/350
Analyzer
Operating Manual, 11/2007
6-271
E-1147.2759.00
Selecting the Difference Frequency Distortion and Measurement Standards
Use
You can select the display mode for the difference frequency distortion
according to the measurement standards:
d2 (IEC 268)
Measurement and display of the 2nd order intermodulation product
acc. to IEC 268
d3 (IEC 268)
Measurement and display of the 3rd order intermodulation product
acc. to IEC 268
d2 (IEC 118)
Measurement and display of the 2nd order intermodulation product
acc. to IEC 118
d3 (IEC118)
Measurement and display of the 3rd order intermodulation product
acc. to IEC 118
Note:
The intermodulation distortion (IMD) tones are not subjected to a level
testing case of measurement according to IEC 118. This means that a DFD
measurement is possible even if the IMD signal is severely distorted (e.g. as
a result of the frequency response of the DUT or of the transmission path). A
typical example is the measurement of earpieces.
Note:
The setting is always valid for both channels (Ch 1&2).
Selecting
measurement
standards
1.
Press the
function key
in the
menu.
A selection field containing the available settings is displayed. The
default setting is “d2 (IEC 268)“.
2.
Use the
rotary knob
[11] to select a setting.
3.
Press the
ENTER key
[5] to close the selection field.
The new setting is stored and displayed in the parameter field.