Selecting the Spacing of Sweep Steps (Lin/Log)
Use
The spacing of the sweep points can be selected. You can decide if the dis-
tance between the valus set one after another is in
Linear
or
Logarithmic
steps.
Selecting the
Spacing
1.
Press the
function key
in the corresponding
or
submenu
.
A selection field containing the available settings is displayed. The
default setting is “Linear“.
2.
Use the
rotary knob
[11] to select a setting.
3.
Press the
ENTER key
[5] to close the selection field.
The new setting is stored and displayed in the parameter field.
Entering the Frequency Resolution
Use
You can enter the frequency resolution in two different ways:
POINTS
Enter the number of reading points.
Based on the frequency range, the generator calculates the position of
the reading points and automatically determines the step size.
STEP SIZE
Enter the step size.
Based on the frequency range, the generator calculates the number of
reading points and automatically determines their position.
The first possibility:
Entering the number
of reading points
1.
Press the
function key
in the
submenu
.
An entry field containing the currently applicable setting is displayed. The
default setting is “100“.
2.
Enter a new value (
Ê
The permissible entry range depends on the frequency range; with
maximum frequency range it is:
2
≤
POINTS
≤
1024
The new setting is stored and displayed in the parameter field.