Multi-purpose connector. The [TRIG] connector is used mainly as an input connector
for an external pulse modulator source.
[I, Q IN / OUT]
The [I IN] and [Q IN] SMA female type connectors are the inputs of the I/Q modulator,
provided for feeding of external signal.
The [I OUT] and [Q OUT] SMA female type connectors are for direct output of the ana-
log I/Q signal.
NOTICE!
Maximum input levels. Do not overload the I and Q inputs. The maximum
permissible voltage is 1V. For details, refer to the data sheet.
Note:
The I/Q modulator requires the hardware optionR&S
SGU-B120V/-B140V.
[AC supply and power switch]
The AC supply and power switch allow you to connect the R&S
SGU to the power sup-
ply and switch on the instrument.