The scan function finds out the connected R&S SGU.
5. Select the R&S SGU from the list and set "Available Instruments > Select > On" to
enable it as an extension.
A green status indicator "Connected" indicates the successfully established remote
connection to the extension.
6. Select "Test Signal Connections > Test" to trigger a check of all required signal
connections.
The diagram displays the connection state of the tested connections.
7. Select "SGMA-GUI > R&S SGS > Freq = 20 GHz", "Lev = -30 dBm" and enable
"RF > State > On".
The extension adopts these values and states automatically. Generated is an I/Q
signal with "RF = 20 GHz" and "Level = -30 dBm".
9.2
How to Restore the LAN Connection to an Instrument
► If the LAN connection to an instrument configured to use a static IP address is lost,
press the [LAN LED] on the instrument's front panel for more than 3 seconds to
reset the LAN settings and to set the