Remote Control
R&S
®
RTM20xx
389
User Manual 1317.4726.02 ─ 01
BUS<b>:I2C:CLOCk:SOURce
<Source>
Sets the input channel to which the clock line is connected.
Suffix:
<b>
.
1..4
Selects the bus.
Parameters:
<Source>
CH1 | CH2 | CH3 | CH4 | D0..D15
CH3 and CH4 are only available with 4-channel R&S
RTM oscil-
loscopes.
*RST:
CH1
BUS<b>:I2C:DATA:SOURce
<Source>
Sets the input channel to which the data line is connected.
Suffix:
<b>
.
1..4
Selects the bus.
Parameters:
<Source>
CH1 | CH2 | CH3 | CH4 | D0..D15
CH3 and CH4 are only available with 4-channel R&S
RTM oscil-
loscopes.
*RST:
CH1
CHANnel<m>:THReshold:FINDlevel
Executes the analysis of all analog channels that are configured for the selected bus and
sets the threshold for digitization of analog signals for each channel.
Suffix:
<m>
.
Selects the input channel. The number of channels depends on
the instrument.
Usage:
Event
I²C - Trigger
................................................................................................391
.................................................................................................391
.................................................................................................391
.................................................................................................392
Command Reference