Remote Control
R&S
®
RTM20xx
276
User Manual 1317.4726.02 ─ 01
Usage:
Query only
*PSC
<Action>
Power on Status Clear
Determines whether the contents of the
ENABle
registers are preserved or reset when
the instrument is switched on. Thus a service request can be triggered when the instru-
ment is switched on, if the status registers ESE and SRE are suitably configured. The
query reads out the contents of the "power-on-status-clear" flag.
Parameters:
<Action>
0 | 1
0
The contents of the status registers are preserved.
1
Resets the status registers.
*RST
ReSeT
Sets the instrument to a defined default status. The default settings are indicated in the
description of commands.
Usage:
Setting only
*SRE
<Contents>
Service Request Enable
Sets the service request enable register to the indicated value. This command determines
under which conditions a service request is triggered.
Parameters:
<Contents>
Contents of the service request enable register in decimal form.
Bit 6 (MSS mask bit) is always 0.
Range:
0 to 255
*STB?
STatus Byte query
Reads the contents of the status byte in decimal form.
Usage:
Query only
*TRG
TRiGger
Command Reference