Search
R&S
®
RTM20xx
138
User Manual 1317.4726.02 ─ 01
"Both"
Rising edge and falling edge
Remote command:
Level
Sets the voltage level for the search.
Remote command:
Hysteresis
Sets a hysteresis range to the search level in order to avoid unwanted search results
caused by noise oscillation around the level. For a rising edge, the hysteresis is below
the search level. Otherwise, for a falling edge the hysteresis is above the level.
Remote command:
10.3.3 Width Setup
Access: SEARCH MENU > "Search type = Width" > "Setup"
Polarity
Indicates the polarity of the pulse to be searched for.
Remote command:
Level
Sets the voltage level on which the pulse width is measured.
Remote command:
Hysteresis
Sets a hysteresis range to the search level in order to avoid unwanted search results
caused by noise oscillation around the level. For a rising edge, the hysteresis is below
the search level. Otherwise, for a falling edge the hysteresis is above the level.
Remote command:
Comparison
Sets the condition how the measured pulse width is compared with the given limit(s).
"Greater than"
Finds pulses longer than the given "Width".
See also:
Reference for Search Menu