![Rohde & Schwarz RT-ZS60 User Manual Download Page 41](http://html.mh-extra.com/html/rohde-and-schwarz/rt-zs60/rt-zs60_user-manual_1477877041.webp)
Putting into Operation
R&S
®
RT-ZS60
19
User Manual 1418.7342.02 ─ 01
Alternatively, Auto Zero can be started in the dialog box of the R&S
RTO:
► "Vertical" menu > "Probe Setup" > "Ch" tab > "Auto zero"
2.2
Micro Button
The micro button provides easy and quick access to important functions of the
R&S
RTO. After a function has been assigned, pressing the micro button remotely
controls this specific function on the base unit. For example, Continuous Run and
Auto Zero are commonly assigned to the micro button.
To assign a function to the micro button with the R&S RTO oscilloscope:
► "Vertical" menu > "Probe Setup" > "Ch" tab > "Micro button action"
For more details on the available functions and settings for the micro button,
refer to the "User Manual" of the R&S oscilloscope.
2.3
Offset Compensation
The Offset Compensation function can compensate a DC component of the input
signal, even in front of the active amplifier in the probe tip. As a result, the entire
dynamic range of the probe is maintained. This function is useful when measuring
AC signals with high superimposed DC component.
Micro Button