![Rohde & Schwarz R&S UPP User Manual Download Page 690](http://html.mh-extra.com/html/rohde-and-schwarz/rands-upp/rands-upp_user-manual_1477869690.webp)
Remote Control - Fundamentals
R&S
®
UPP
674
User Manual 1411.1055.32 ─ 10
Bit No. Meaning
1
Ch2
0
Ch1
6.8.10
STATus:QUEStionable:MEASuring Register
This register only contains information which channels are switched on or off for multi-
channel operation. It can be read by the query
STATus:QUEStionable:MEASuring:CONDition?
or
STATus:QUEStionable:MEASuring[:EVENt]?
.
Reading the EVENt part deletes its content!
Table 6-8: STATus:QUEStionable:MEASuring Register
Bit No. Meaning
15
Slave 8
For cascading, signals with a '1' which slave has identified a change of the channel
setting.
14
Slave 7
13
Slave 6
12
Slave 5
11
Slave 4
10
Slave 3
9
Slave 2
8
Slave 1
7
Ch8
Returns the setting of channels of multichannel operation in the Analyzer Config
panel.
0 = The channel in question is deactivated
1 = The channel in question is activated
The associated summary bit is in the STATus:QUEStionable register bit 0
6
Ch7
5
Ch6
4
Ch5
3
Ch4
2
Ch3
1
Ch2
0
Ch1
If a two-channel analyzer is switched on, the bits 0 ... 15 are deleted.
Status Reporting System
Summary of Contents for R&S UPP
Page 1: ...R S UPP Audio Analyzer User Manual User Manual Version 10 1411105532 ÅP ...
Page 34: ...Contents R S UPP 18 User Manual 1411 1055 32 10 ...
Page 630: ...Instrument Functions R S UPP 614 User Manual 1411 1055 32 10 Format ...
Page 631: ...Instrument Functions R S UPP 615 User Manual 1411 1055 32 10 Format ...