Instrument Functions
R&S
®
UPP
441
User Manual 1411.1055.32 ─ 10
Click the Show button to open the "Dig
Analyzer
Protocol" window. You can close the
window, as with any other, by means of clicking on the title bar button .
"deactivated"
The protocol analyzer is off.
This setting is recommended for increasing the measurement speed
(e.g. in the case of sweeps or in remote-control mode), if protocol
data is not currently needed.
The current channel-status information is frozen when this feature is
switched off.
"activated"
The protocol analyzer is on. The parameter fields shown above are
displayed.
Note: Protocol analysis in the "Digital Audio" analyzer instru-
ment
:
The internal hardware determines the channel-status information for
channels 1 and 2 on a sequential basis. This information is therefore
available only after 2*192
samples at the earliest. A single measure-
ment must therefore capture at least 384
samples, which can
lengthen the measurement time during fast measurements.
Protocol analysis in the "HDMI" analyzer instrument:
Since the stream header has no channel reference for HDMI, the
channels are always represented as identical in the protocol analyzer.
Remote command:
ON | OFF
Display Mode
For selecting the display format of the channel-status information.
Depending on the format bit (bit #0 of the channel-status bits), the following bit fields
have a different meaning. This menu item defines whether the channel-status informa-
tion is interpreted according to the Consumer format or the Professional format.
"Automatic"
The channel-status bits are interpreted in the format specified by bit
#0. If this bit changes, the format is changed automatically to the cor-
rect one.
Protocol Analysis
Summary of Contents for R&S UPP
Page 1: ...R S UPP Audio Analyzer User Manual User Manual Version 10 1411105532 ÅP ...
Page 34: ...Contents R S UPP 18 User Manual 1411 1055 32 10 ...
Page 630: ...Instrument Functions R S UPP 614 User Manual 1411 1055 32 10 Format ...
Page 631: ...Instrument Functions R S UPP 615 User Manual 1411 1055 32 10 Format ...