Instrument Functions
R&S
®
UPP
313
User Manual 1411.1055.32 ─ 10
"Auto"
The sampling rate is measured, set and displayed in the following
line. If the applied sampling rate is changed the set value is automati-
cally updated.
Note:
The digital analyzer must re-initialized for each new sampling
frequency. To avoid the need to repeat this time-consuming proce-
dure too often, minor deviations and fluctuations (below 3
Hz) are
ignored.
"32 kHz to
192 kHz"
Standard sampling rate 32 kHz to 192 kHz
"Value "
Variable input of sampling rate
Remote command:
INPut[1]:SAMPle:FREQuency:MODE
Pre Filter
The selected prefilter affects all measurement functions. This permits desired prefilter-
ing without selecting appropriate filters for the individual measuring functions.
Remote command:
Input
Selection of the input interface for the Digital Audio Analyzer with the R&S
UPP-B2
(digital audio interfaces) option
"Unbal (BNC)"
Unbalanced digital audio input at the BNC socket
"Optical"
Optical digital audio input (TOSLINK).
"Bal (D-Sub)"
Balanced digital audio input at the D-Sub male connector on the front
panel
"Intern"
Internal connection to the Digital Audio Generator
Loop measurement possible without connected cable.
Remote command:
Start Cond
Specifies the event which triggers a measurement.
Configuring the Digital Analyzer
Summary of Contents for R&S UPP
Page 1: ...R S UPP Audio Analyzer User Manual User Manual Version 10 1411105532 ÅP ...
Page 34: ...Contents R S UPP 18 User Manual 1411 1055 32 10 ...
Page 630: ...Instrument Functions R S UPP 614 User Manual 1411 1055 32 10 Format ...
Page 631: ...Instrument Functions R S UPP 615 User Manual 1411 1055 32 10 Format ...