Instrument Functions
R&S
®
UPP
250
User Manual 1411.1055.32 ─ 10
Several thousand equidistant sine-wave lines in a selectable frequency domain
Special application:
If the frequency grid of the sine-wave lines generated here is identical to the analysis
grid of the FFT, analysis can be performed without leakage in the FFT. It is therefore
possible to obtain a selectivity of one line with the square
‑
wave window. This type of
generator/analyzer configuration allows frequency responses of DUTs to be deter-
mined in one go with high precision and selectivity.
Function
Selects the random noise signal.
Remote command:
RANDom
Shape
Defines the amplitude distribution of the signal spectrum within a selectable frequency
domain.
"White"
Time domain:
Uncorrelated white noise without periodicity is produced.
Frequency range:
Uncorrelated white noise without periodicity is produced.
Generator Signals