Instrument Functions
R&S
®
UPP
248
User Manual 1411.1055.32 ─ 10
Dither
In the digital generator, a noise component with adjustable amplitude can be superim-
posed on the signal. The checkbox enables or disables this function.
The parameter fields for Dither do not appear in the analog generator.
Remote command:
Sweep Ctrl
Used to activate the sweep system. The center frequency (in the IEC 118 mode) or the
upper frequency (in the IEC 268 mode) or the total level (Total Voltage) of the signal
can be varied from the start to the stop value. If frequency or level is selected as the x-
axis, the measured values can be displayed as a curve in the Sweep Graph graphical
window.
Sweep points: Linear or logarithmic between Start and Stop or read out of the file.
Forwarding: Automatic in a defined timeframe or synchronized to the measurement
time of the analyzer. All setting parameters of the sweep system are described in detail
in
Remote command:
Mean Freq
Used to set the center frequency (in the IEC 268 mode).
Remote command:
Upper Freq
Used to set the upper frequency (in the IEC 118 mode).
Remote command:
Diff Freq
Used to set the frequency difference of the two sine waves.
Remote command:
Total Voltage
Used to set the level of the overall signal. The maximum value depends on the config-
ured DC offset. In the analog generator, the maximum total voltage also depends on
the settings
on page 199 in the Gener-
ator Config panel.
Remote command:
Generator Signals
Summary of Contents for R&S UPP
Page 1: ...R S UPP Audio Analyzer User Manual User Manual Version 10 1411105532 ÅP ...
Page 34: ...Contents R S UPP 18 User Manual 1411 1055 32 10 ...
Page 630: ...Instrument Functions R S UPP 614 User Manual 1411 1055 32 10 Format ...
Page 631: ...Instrument Functions R S UPP 615 User Manual 1411 1055 32 10 Format ...