Instrument Functions
R&S
®
UPP
215
User Manual 1411.1055.32 ─ 10
The sampling frequency corresponds to the frequency of the FSYNC signal. All fre-
quency information of generated audio data relates to this setting. If the I
2
S generator
is operated with external wordclock synchronization and the frequency of the external
clock signal does not correspond to the set sampling frequency, the I
2
S generator gen-
erates audio data with an incorrect frequency. Even with external masterclock synchro-
nization, the external clock signal must correspond to the set sampling frequency multi-
plied by the selected MClk ratio. Otherwise, the I
2
S generator also generates audio
data with an incorrect frequency.
The selected clock rate determines the maximum signal frequency which can be input.
For sine-wave signals, this is 49.9
% of the sampling rate.
The standard sampling frequencies can be set directly in the selection field.
The selection item "Value" permits the variable input of the sampling rate in the follow-
ing line.
Value range for numeric input:
6.75
kHz to 200
kHz
Remote command:
MClk Freq
Display of the master clock frequency based on the relationship: "Sample Freq" *
"MClk Ratio".
Value range: 864
kHz to 49.152
MHz
Remote command:
BClk Freq
Display of the bit clock frequency. This is calculated from the product of Word Length /
Sample Freq.
Value range: 216
kHz to 12.80
MHz
Configuring the I
2
S Generator
Summary of Contents for R&S UPP
Page 1: ...R S UPP Audio Analyzer User Manual User Manual Version 10 1411105532 ÅP ...
Page 34: ...Contents R S UPP 18 User Manual 1411 1055 32 10 ...
Page 630: ...Instrument Functions R S UPP 614 User Manual 1411 1055 32 10 Format ...
Page 631: ...Instrument Functions R S UPP 615 User Manual 1411 1055 32 10 Format ...