Remote Control Commands
R&S
®
RTM3000
499
User Manual 1335.9090.02 ─ 08
ACQuire:COUNt
Sets the number of segments to be captured.
In automatic mode, the command is a query and returns the maximum number of seg-
ments that can be captured with the current configuration. See also:
Parameters:
<NoOfAcquisitions> Number of available segments in the memory
ACQuire:NSINgle:COUNt
<NSingleCount>
Sets the number of waveforms acquired with
Parameters:
<NSingleCount>
Number of waveforms.
Range:
1 to maximum number that depends on the record
length.
*RST:
1
ACQuire:AVAilable?
Returns the number of segments that are currently saved in the memory. This number
is available for history viewing.
Parameters:
<Acquisitions>
Number of captured segments
Usage:
Query only
ACQuire:SEGMented:STATe
<State>
If ON, the acquisitions are performed as fast as possible, without processing and dis-
playing the waveforms. When acquisition has been stopped, the data is processed and
the latest waveform is displayed. Older waveforms are stored in segments. You can
display and analyze the segments using the history..
Parameters:
<State>
ON | OFF
ACQuire:AVERage:CURRent?
Returns the current amount of acquired waveforms that contribute to the average. The
value is independent of the number of available segments, there are more waveforms
in the average available than segments in history.
Parameters:
<CurrentAverages>
Range:
Minimum is 2; maximum depends on the instrument
settings.
Usage:
Query only