Instrument Functions
R&S
®
NGL200/NGM200
58
User Manual 1178.8736.02 ─ 09
6.5
Protection
There are various ways in which the R&S
NGL/NGM protects itself and the connected
load from damage due to overvoltage, overcurrent and overpower drawn by the load
during testing.
1. Press [Settings] key.
The R&S
NGL/NGM displays the device/channel menu window.
2. Select the required channel tab to configure the various protection dialogs.
The R&S
NGL/NGM displays the selected channel menu.
6.5.1
Overcurrent Protection (OCP)
When the drawn current exceeds the limit set for the respective channels, an alert is
triggered and the affected channels are turned off according to the settings configured
in the OCP dialog.
1. Select "Overcurrent Protection (OCP)" from the menu.
The R&S
NGL/NGM displays the OCP dialog.
Protection