Remote Control Commands
R&S
®
NGL200/NGM200
109
User Manual 1178.8736.02 ─ 09
0
Deactivates the safety limit.
Parameters for setting and query:
<Channel list>
<list>
Example:
ALIM 1, (@1)
Activates the safety limit state at channel 1
Example:
See
Example "Configuring the output voltage"
[SOURce:]VOLTage[:LEVel][:IMMediate]:ALIMit:LOWer
<voltage>
[SOURce:]VOLTage[:LEVel][:IMMediate]:ALIMit:LOWer?
[<Channel list>]
Sets or queries the lower safety limit for voltage.
Setting parameters:
<voltage>
<numeric value> | MIN | MINimum | MAX | MAXimum | <list>
<numeric vaule>
Numeric value for safety limit.
MIN | MINimum
Min value for lower safety limit.
MAX | MAXimum
Max value for lower safety limit.
Range:
0.000E+00 to 2.050E+01
Increment: 0.001
*RST:
0.000E+00
Parameters for setting and query:
<Channel list>
<list>
Example:
VOLT:ALIM:LOW? (@1)
Queries the lower safety limit for volatage at channel 1
Example:
See
Example "Configuring the output voltage"
[SOURce:]VOLTage[:LEVel][:IMMediate]:ALIMit[:UPPer]
<voltage>
[SOURce:]VOLTage[:LEVel][:IMMediate]:ALIMit[:UPPer]?
[<Channel list>]
Sets or queries the upper safety limit for voltage.
Setting parameters:
<voltage>
<numeric value> | MIN | MINimum | MAX | MAXimum | <list>
<numeric value>
Numeric value for upper safety limit.
MIN | MINimum
Min value for upper safety limit.
MAX | MAXimum
Max value for upper safety limit.
Range:
0.000E+00 to 2.050E+01
Increment: 0.001
Configuration Commands