Instrument functions
R&S
®
LCX Series
94
User Manual 1179.2260.02 ─ 02
The R&S
LCX uploads the data and displays all information on the screen.
Figure 8-19: Log chart result
The labels of the parameter softkeys indicate the measured parameter with the
respective unit per grid in the diagram. The colored bar of each softkey corre-
sponds to the trace color.
8.8.3
Logging
The logging function enables you to record the measurement data. You can save the
readings on the internal memory of the R&S
LCX, or on an externally connected mem-
ory device, e.g. a USB flash drive. The function saves the data in a
.csv
file with
generically generated file name.
Aspects on the logging function:
●
The R&S
LCX does not support external USB hard drives (or USB extensions).
We recommend that you use USB flash drives with FAT/FAT32 formatting. The
instrument recognizes the USB device automatically.
●
The duration for capturing logging data in the range from 0
s to 3.49*10
5
s at a
maximum.
●
When running, the instrument indicates a white "LOG" label in the status bar. When
disabled, the label color is gray.
8.8.3.1
Logging settings
In the "Logging" menu, you can configure the main parameter settings for data logging.
With a USB stick connected, you can save the file on either, the USB stick or the inter-
nal memory, with "USB1A" as default setting. Without USB connected, the R&S
LCX
saves the file internally in the file path
/int/logging
.
Access:
1. Press the [settings] key at the front panel.
Specific instrument functions