Measurement setups
R&S
®
LCX Series
54
User Manual 1179.2260.02 ─ 02
6.1.3
Test fixture for SMD components
The R&S
LCX-Z3 test fixture is designed for measuring SMD, i.e. leadless compo-
nents. It is suitable for rectangular or cylindrical shaped chips with parallel electrodes,
and components similar in shape and type.
Figure 6-3: R&S
LCX-Z3 SMD test fixture
The test fixture provides a section to position and center the component and clamp it
between two connection pins.
To perform short correction, close the contacts of the connection pins. For open correc-
tion, use the lever to open the pins until the spacing between them corresponds
approximately to the length of the measurement sample. Hold this position during the
correction.
6.1.4
Test tweezers for SMD components
The test tweezers R&S
LCX-Z4) test fixture is designed for measuring SMD compo-
nents, even when they are mounted on printed circuit boards. It provides simple and
fast contacting and is thus suitable, for example, for pre-selection during incoming
inspection or of small batches.
About test fixtures
www.allice.de
Allice Messtechnik GmbH