Measurement basics
R&S
®
LCX Series
50
User Manual 1179.2260.02 ─ 02
Parameter
Designation and unit
Lp-Rp
parallel inductance (H) - resistance (
Ω
)
Lp-G
parallel inductance (H) - conductance
Rs-X
*)
series resistance (
Ω
) - reactance (
Ω
)
G-B
*)
conductance (S) - susceptance (
Ω
)
Z-
Θ
r
*)
impedance (
Ω
) - phase angle (rad)
Z-
Θ
d
*)
impedance (
Ω
) - phase angle (°)
Y-
Θ
r
*)
admittance (S) - phase angle (rad)
Y-
Θ
d
*)
admittance (S) - phase angle (°)
*)
selectable as second measurement parameter pair
Table 5-5: Transformer measurement parameters
Parameter
Designation and unit
M-
Θ
d
mutual inductance (H) - phase angle (°)
N-
Θ
d
transformer ratio - phase angle (°) (phase difference between primary
and secondary winding)
Rs-X
*)
series resistance (
Ω
) - reactance (
Ω
)
G-B
*)
conductance (S) - susceptance (
Ω
)
Z-
Θ
r
*)
impedance (
Ω
) - phase angle (rad)
Z-
Θ
d
*)
impedance (
Ω
) - phase angle (°)
Y-
Θ
r
*)
admittance (S) - phase angle (rad)
Y-
Θ
d
*)
admittance (S) - phase angle (°)
*)
selectable as second measurement parameter pair
5.3
Considerations on measurement accuracy
To know the measurement accuracy of a, e.g. capacitance measurement, the accuracy
and characteristics of the instrument are essential.
An R&S
LCX LCR meter measures the impedance (Z) and phase (
Θ
), and derives the
results of the also provided measurement parameters. The accuracy of a determined
parameter therefore refers to the accuracy of these measured parameters.
During a measurement, the R&S
LCX displays the accuracy of (Z) and (
Θ
) at the cur-
on the screen.
If you need to know the accuracy outside a running measurement or for a secondary
measurement parameter, you can calculate the accuracy individually.
Considerations on measurement accuracy