Remote control commands
R&S
®
LCX Series
201
User Manual 1179.2260.02 ─ 02
Example: Logging data settings
The example represents the remote control commands for configuring the logging
mode.
****************************************
// Configure the logging function.
// Deactivate logging mode to set the parameters:
****************************************
:LOG:STATe OFF
****************************************
// Configure the parameters:
// select 10 min time interval, SPAN mode, start time and duration.
:LOG:INTerval 300
:LOG:MODE SPAN
:LOG:STIMe 2020,5,1,13,30,00
:LOG:DURation 150
// :LOG:MODE COUNt
// :LOG:COUNt 100
****************************************
// Query the data logging settings
:LOG:MODE?
// Response: SPAN
:LOG:FNAME?
// Response: 'log-20210225T105014.455.csv', EXT
:LOG:INTerval?
/// Response 300
:LOG:STIMe?
// Response: 2020,5,1,13,30,00 // 2020-05-01, 13:30 pm
:LOG:DURation?
// Response: 150 // 2 h 30 min,0 s
:LOG:COUNt?
// Response: 100
:LOG:COUNt? MIN
// Response: 1
:LOG:COUNt? MAX
// Response: 1000000
****************************************
// Start and stop logging data
:LOG[:STATe] ON
:LOG[:STATe]?
// Response: 1 // logging data is activated
****************************************
// Restart logging of a new measurement
LOG subsystem