![Rohde & Schwarz R&S LCX Series User Manual Download Page 146](http://html.mh-extra.com/html/rohde-and-schwarz/rands-lcx-series/rands-lcx-series_user-manual_1477867146.webp)
Network operation and remote control
R&S
®
LCX Series
146
User Manual 1179.2260.02 ─ 02
GPIB address
The controller addresses the instrument with the GPIB bus channel, see
on page 117. GPIB provides channel addresses from 0 to 30. The
GPIB address is preset to 20 (factory-set), but you can adjust it according to your net-
work environment, see
"Assigning the GPIB address manually"
ment preset does not affect the GPIB address.
The GPIB resource string syntax is as follows:
GPIB:://<channel>[::INSTR]
GPIB
= denotes the used interface
<channel address> = the used channel
[::INSTR]
= indicates the instrument resource class (optional)
Note:
If the VISA implementation supports the GPIB interface, you can optionally
define the VISA Instrument Control Resource (INSTR). It is used to define the basic
operations and attributes for a device, such as reading, writing, or triggering.
Example:
GPIB::28
28 is the selected GPIB bus channel
10.2.4
Status reporting system
The status reporting system saves all information on the current operating state of the
instrument and errors which have occurred. This information is saved in the status reg-
isters and in the error queue. You can query the status of the registers remotely, i.e.
over the USB, GPIB or LAN interfaces with the commands of the
Remote control interfaces and protocols
www.allice.de
Allice Messtechnik GmbH