4.8
Operating Manual 1164.4556.12 - 06
4.4 Return to Manual Operation – LOCAL Menu
LOCAL
The
LOCAL
menu is displayed on switching the instrument to remote control mode.
At the same time, the hotkey bar is blanked out and all keys are disabled except the
PRESET
key.
The
LOCAL
softkey and the
DISPLAY UPDATE ON/OFF
softkey are displayed.
Depending on the setting of the
DISPLAY UPDATE ON/OFF
softkey, the diagrams,
traces and diplay fields are displayed or hidden. For further details on the
DISPLAY
UPDATE ON/OFF
Instrument Setup and Interface Configuration –
The
LOCAL
key switches the instrument from remote to manual operation, with the
assumption that the remote controller has not previously set the LOCAL LOCKOUT
function.
A change in the control mode consists of:
•
Enabling the Front Panel Keys
Returning to manual operation enables all inactive keys and turns on the hotkey
bar. The softkey menu which is displayed is the main menu of the current mode.
•
Inserting the measurement diagrams
The blanked diagrams, traces and display fields are inserted.
•
Generating the message OPERATION COMPLETE
If, at the time of pressing the
LOCAL
softkey, the synchronization mechanism via
*OPC, *OPC? or *WAI is active, the currently running measurement procedure is
aborted and synchronization is achieved by setting the corresponding bits in the
registers of the status reporting system.
•
Setting Bit 6 (User Request) of the Event Status Register
With a corresponding configuration of the status reporting system, this bit
immediately causes the generation of a service request (SRQ) which is used to
inform the control software that the user wishes to return to front-panel control.
This information can be used, e.g., to interrupt the control program so that the user
can make necessary manual corrections to instrument settings. This bit is set
each time the
LOCAL
softkey is pressed.
If the LOCAL LOCKOUT function is active in the remote control mode, the front-
panel
PRESET
key is also disabled. The LOCAL LOCKOUT state is left as soon
as the process controller de-activates the REN line or the GPIB cable is discon-
nected from the instrument.
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...