Device-specific messages
-420
Query UNTERMINATED
The query is incomplete.
Example: The instrument is addressed as a talker and receives incomplete data.
-430
Query DEADLOCKED
The query cannot be processed.
Example: The input and output buffers are full, the instrument cannot continue
operation.
-440
Query UNTERMINATED after indefinite response
A query is in the same command line after a query which requests an indefinite
response.
Error code
Error text in the case of queue poll
Error explanation
Error code
Error text in the case of queue poll
Error explanation
1036
MS: The correction table based amplifier gain exceeds the amplifier range for
CALAMP1 and CALAMP2 on IF board
This message is output when the setting range of the calibration amplifier is not
sufficient for the required correction. The error occurs only with modules which are not
correctly adjusted or defective.
1052
Frontend LO is Unlocked
This message is output when the phase regulation of the local oscillator fails in the RF
frontend.
1060
Trigger-Block Gate Delay Error- gate length < Gate Delay
This message is output when the gate signal length is not sufficient for the pull-in delay
with a predefined gate delay.
2022
OPTIONS.INI invalid
This message is output when an error has been recognized in the file OPTIONS.INI
which contains the clearing codes for retrofitable firmware applications. If this file is not
correctly recognized, all firmware applications are blocked for this instrument.
2028
Hardcopy not possible during measurement sequence
This message is output when a printout is started during scan sequences that cannot
be interrupted. Such sequences are for example:
• Recording the system error correction data (calibration)
• Instrument selftest
In such cases a synchronization to the end of the scan sequence should be performed
prior to starting the printout.
2033
Printer Not Available
This message is output when the selected printer is not included in the list of available
output devices. A possible cause is that the required printer driver is missing or
incorrectly installed.
2034
CPU Temperature is too high
This message is output when the temperature of the processor exceeds 70
°
C.
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...