R&S FSP
Remote Control – Description of Commands
STATus Subsystem
6.220
Operating Manual 1164.4556.12 - 06
STATus:QUEStionable:POWer:PTRansition
0 to 65535
This command sets the edge detectors of all bits of the STATus:QUEStionable:POWer register from 0
to 1 for the transitions of the CONDition bit.
Example:
"STAT:QUES:POW:PTR 65535"
Characteristics:
*RST value: –
SCPI: conform
Mode:
A
STATus:QUEStionable:PTRansition
0 to 65535
This command sets the edge detectors of all bits of the STATus:QUEStionable register from 0 to 1 for
the transitions of the CONDition bit.
Example:
"STAT:QUES:PTR 65535"
Characteristics:
*RST value: –
SCPI: conform
Mode:
A
STATus:QUEue[:NEXT]?
This command returns the earliest entry to the error queue and deletes it.
Positive error numbers indicate device-specific errors, negative error numbers are error messages
defined by SCPI (cf. chapter
). If the error queue is empty, the error number 0, "no
error", is returned. This command is identical with the command
SYSTem:ERRor
.
Example:
"STAT:QUE?"
Characteristics:
*RST value: –
SCPI: conform
Mode:
A
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...