Operating Manual 1164.4556.12 - 06
6.119
R&S FSP
Remote Control – Description of Commands
HCOPy Subsystem
HCOPy:DEVice:LANGuage<1|2>
GDI | WMF | EWMF | BMP
This command determines the data format for the printed output.
Parameter:
GDI (Graphics Device Interface): Default format for output to a printer that has
been configured under Windows. Must be selected for output to the printer
interface (
HCOPy:DEVice 'SYST:COMM:PRIN'
). Can be used for output to a file
(
HCOPy:DEVice 'SYST:COMM:MMEM'
). The printer driver that was configured
under Windows is used to generate a file format that is specific to the printer. GDI
is available only for HCOPY:MODE SCReen.
WMF (WINDOWS Metafile) and EWMF (Enhanced Metafile Format): Data formats
for output to files which can be directly processed at a later point in time for
documentation purposes using suitable software. WMF can be used only for output
to a file (
HCOPy:DEVice 'SYST:COMM:MMEM'
), EWMF can also be used for
output to the clipboard (
HCOPy:DEVice 'SYST:COMM:CLIP'
).
BMP (Bitmap): Data format, exclusively for output to files (
HCOPy:DEVice
'SYST:COMM:MMEM'
).
BMP is available only for HCOPY:MODE SCReen.
Example:
"HCOP:DEV:LANG WMF"
Characteristics:
*RST value: -
SCPI: conform
Mode:
A
HCOPy[:IMMediate<1|2>]
This command starts a hardcopy output. The numeric suffix selects which printer configuration (1 or 2)
is to be used for the hardcopy output. If there is no suffix, configuration 1 is automatically selected.
Example:
"HCOP"
"HCOPy:IMM1"
'Starts the hardcopy output to device 1 (default).
"HCOPy:IMM2"
'Starts the output to device 2.
Characteristics:
*RST value: -
SCPI: conform
Mode:
A
This command is an event and therefore has no *RST value and no query.
HCOPy:ITEM:ALL
This command selects the complete screen to be output.
The hardcopy output is always provided with comments, title, time and date. As an alternative to the
whole screen, only traces (commands
'HCOPy:ITEM:WINDow:TRACe:STATe ON'
) or tables
(command
'HCOPy:ITEM:WINDow:TABLe:STATe ON'
) can be output.
Example:
"HCOP:ITEM:ALL"
Characteristics:
*RST value: -
SCPI: conform
Mode:
A
This command is an event and therefore has no *RST value and no query.
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...