R&S FSP
Remote Control – Description of Commands
DISPlay Subsystem
6.112
Operating Manual 1164.4556.12 - 06
DISPlay[:WINDow<1|2>]:TRACe<1...3>:X:SPACing
LINear | LOGarithmic
This command toggles between linear and logarithmic display.
Example:
"DISP:TRAC:X:SPAC LIN"
Characteristics:
*RST value: LOGarithmic
SCPI: conform
Mode:
A
The numeric suffix in TRACe<1...3> is irrelevant.
DISPlay[:WINDow<1|2>]:TRACe<1...3>:Y[:SCALe]
10dB to 200dB
This command defines the display range of the y-axis (level axis) in the selected measurement window
with logarithmic scaling (
DISPlay[:WINDow<1|2>]:TRACe<1...3>:Y[:SCALe]:RVALue
).
For linear scaling, the display range is fixed and cannot be modified. The numeric suffix at
TRACe<1...3> is irrelevant.
Example:
"DISP:TRAC:Y 110dB"
Characteristics:
*RST value: 100dB
SCPI: device-specific
Mode:
A
DISPlay[:WINDow<1|2>]:TRACe<1...3>:Y[:SCALe]:MODE
ABSolute | RELative
This command defines the scale type of the y-axis (absolute or relative) in the selected measurement
window.
SYSTem:DISPlay
is set to
OFF,
this command has no immediate effect on the screen. The
numeric suffix in TRACe<1...3> is irrelevant.
Example:
"DISP:TRAC:Y:MODE REL"
Characteristics:
*RST value: ABS
SCPI: device-specific
Mode:
A
DISPlay[:WINDow<1|2>]:TRACe<1...3>:Y[:SCALe]:RLEVel
-
130dBm to 30dBm
This command defines the reference level in the selected measurement window. Depending on the
coupling of the measurement windows, it is valid for both screens or only for the selected measurement
window (
). The unit depends on the setting defined with the
command.
The numeric suffix at TRACe<1...3> is irrelevant.
If the reference level offset is not 0 (
DISPlay[:WINDow<1|2>]:TRACe<1...3>:Y[:SCALe]:RLEVel:
), the indicated value range of the reference level is modified by the offset.
Example:
"DISP:TRAC:Y:RLEV -60"
Characteristics:
*RST value: -20
SCPI: conform
Mode:
A
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...