Measurements on Modulated Signals
Fig. 2-2
Measuring the AF signal from a 1 kHz AM carrier
If the R&S FSP is equipped with the AM/FM Demodulator option (R&S FSP-
B3), the AF can be monitored on the built-in loudspeaker.
5. Switch on the internal AM demodulator
➢
Press the
MKR FCTN
key.
➢
Press the
MKR DEMOD
softkey.
The R&S FSP switches the AM demodulator on automatically.
➢
Turn up volume control.
A 1 kHz tone is output by the built-in loudspeaker.
2.5.1.2
Measurement Example 2 – Measuring the modulation depth of an AM carrier in
the frequency domain
Test setup:
Settings on the signal generator (e.g. R&S SMIQ):
Measurement with the R&S FSP:
1. Set the R&S FSP to its default state
➢
Press the PRESET key.
The R&S FSP is in its default state.
2. Set the center frequency to 100 MHz and the span to 0 kHz
➢
Press the
FREQ
key and enter
100 MHz
.
Frequency:
100 MHz
Level:
-30 dBm
Modulation:
50 % AM, 1 kHz AF