Remote Control – Description of Commands
CALCulate Subsystem
6.94
Operating Manual 1164.4556.12 - 06
CALCulate<1|2>:STATistics:NSAMples
100 to 1E9
This command sets the number of measurement points to be acquired for the statistical measurement
functions.
Example:
"CALC:STAT:NSAM 500"
'Sets the number of measurement points to be acquired to 500.
Characteristics:
*RST value: 100000
SCPI: device-specific
Mode:
A
CALCulate<1|2>:STATistics:PRESet
This command resets the scaling of the X and Y axes in a statistical measurement. The following
values are set:
• x-axis ref level: -20 dBm
• x-axis range APD: 100 dB
• x-axis range CCDF: 20 dB
• y-axis upper limit: 1.0
• y-axis lower limit: 1E-6
Example:
"CALC:STAT:PRES"
'
Resets the scaling for statistical functions
Characteristics:
*RST value: --
SCPI: device-specific
Mode:
A
This command is an event and therefore has no *RST value and no query.
CALCulate<1|2>:STATistics:RESult<1...3>?
MEAN | PEAK | CFACtor | ALL
This command reads out the results of statistical measurements of a recorded trace. The trace is
selected with the numeric suffix <1...3> attached to RESult.
The required result is selected via the following parameters:
Parameter:
MEAN: Average (=RMS) power in dBm measured during the measurement time.
PEAK: Peak power in dBm measured during the measurement time.
CFACtor: Determined CREST factor (= ratio of peak power to average power) in
dB.
ALL: Results of all three measurements mentioned before, separated by commas:
<mean power>,<peak power>,<crest factor>
Example:
"CALC:STAT:RES2? ALL"
'Reads out the three measurement results of trace 2. Example of answer string:
5.56,19.25,13.69 i.e. mean power: 5.56 dBm, peak power 19.25 dBm, CREST
factor 13.69 dB
Characteristics:
*RST value: --
SCPI: device-specific
Mode:
A
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...