Operating Manual 1164.4556.12 - 06
6.87
R&S FSP
Remote Control – Description of Commands
CALCulate Subsystem
CALCulate<1|2>:MARKer<1...4>:FUNCtion:SUMMary:RMS:RESult?
This command queries the result of the measurement of the RMS power value in the selected
measurement window.
The function is independent of the marker selection, i.e. the numeric suffix <1...4> of
:MARKer
is
irrelevant. It is only available in the time domain (span = 0).
A complete sweep with synchronization to sweep end must be performed between switching on the
function and querying the measured value to obtain a valid query result. This is only possible in single-
sweep mode.
Example:
"INIT:CONT OFF"
'Switches to single-sweep mode.
"CALC:MARK:FUNC:SUMM:RMS ON"
'Switches on the function in screen A.
"INIT;*WAI"
'Starts a sweep and waits for the end.
"CALC:MARK:FUNC:SUMM:RMS:RES?"
'Outputs the result of screen A.
Characteristics:
*RST value: -
SCPI: device-specific
Mode:
A-T
This command is only a query and therefore has no *RST value.
CALCulate<1|2>:MARKer<1...4>:FUNCtion:SUMMary:RMS[:STATe]
ON | OFF
This command switches on or off the measurement of the effective (RMS) power in the selected
measurement window. If necessary the function is switched on previously.
The function is independent of the marker selection, i.e. the numeric suffix <1...4> of
:MARKer
is
irrelevant. It is only available in the time domain (span = 0).
Example:
"CALC2:MARK:FUNC:SUM:RMS ON"
'Switches on the function in screen B.
Characteristics:
*RST value: OFF
SCPI: device-specific
Mode:
A-T
CALCulate<1|2>:MARKer<1...4>:FUNCtion:SUMMary:SDEViation:AVERage:RESult?
This command queries the result of the averaged standard deviation determined in several sweeps in
the selected measurement window. The query is possible only if averaging is active. The function is
independent of the marker selection, i.e. the numeric suffix <1...4> of
:MARKer
is irrelevant. It is only
available in the time domain (span = 0).
A complete sweep with synchronization to sweep end must be performed between switching on the
function and querying the measured value to obtain a valid query result. This is only possible in single-
sweep mode.
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...