R&S FSP
Remote Control – Description of Commands
CALCulate Subsystem
6.52
Operating Manual 1164.4556.12 - 06
CALCulate<1|2>:MARKer<1...4>:X:SLIMits:RIGHt
0 to MAX (frequency | sweep time)
This command sets the right limit of the search range for markers and delta markers in the selected
measurement window. Depending on the x-axis domain the indicated value defines a frequency (span
> 0) or time (span = 0). The function is independent of the selection of a marker, i.e. the numeric suffix
in MARKer<1...4> is irrelevant.
If the time domain power measurement is active, this command limits the evaluation range to the trace.
Example:
"CALC:MARK:X:SLIM ON"
'Switches the search limit function on for screen A.
"CALC:MARK:X:SLIM:RIGH 20MHz"
'Sets the right limit of the search range in screen A to 20 MHz.
Characteristics:
*RST value: - (is set to the right diagram border on switching on search limits)
SCPI: device-specific
Mode:
A
CALCulate<1|2>:MARKer<1...4>:X:SLIMits[:STATe]
ON | OFF
This command switches between a limited (ON) and unlimited (OFF) search range in the selected
measurement window. The function is independent of the selection of a marker, i.e. the numeric suffix
MARKer<1...4> is irrelevant.
If the time domain power measurement is active, this command limits the evaluation range on the
trace.
Example:
"CALC:MARK:X:SLIM ON"
'Switches on search limitation in screen A.
Characteristics:
*RST value: OFF
SCPI: device-specific
Mode:
A
CALCulate<1|2>:MARKer<1...4>:Y:PERCent
0 to100%
This command positions the selected marker in the selected window to the given probability. If marker
2, 3 or 4 is selected and used as a delta marker, it is switched to marker mode.
Example:
"CALC1:MARK:Y:PERC 95PCT"
'Positions marker 1 in screen A to a 'probability of 95%.
Characteristics:
*RST value: -
SCPI: device-specific
Mode:
A
The function is only available if the search limit for marker and delta marker is
switched on (
CALC:MARK:X:SLIM ON
).
The command is only available with the CCDF measurement switched on.
The associated level value can be determined with the
CALC:MARK:X?
command.
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...