R&S FSP
Remote Control – Description of Commands
CALCulate Subsystem
6.28
Operating Manual 1164.4556.12 - 06
CALCulate<1|2>:LIMit<1...8>:DELete
This command deletes the selected limit line.
Up to 8 limit lines can be defined at the same time. The command is independent of the measurement
window.
Example:
"CALC:LIM1:DEL"
'Deletes limit line 1.
Characteristics:
*RST value: --
SCPI: device-specific
Mode:
A
This command is an event and therefore has no *RST value and no query.
CALCulate<1|2>:LIMit<1...8>:FAIL?
This command queries the result of the limit check of the limit line. It should be noted that a complete
sweep must have been performed for obtaining a valid result. A synchronization with
*OPC
,
*OPC?
or
*WAI
should therefore be provided.
Up to 8 limit lines can be defined at the same time. The numeric suffixes <1|2> of CALCulate indicate
the measurement window. The result of the limit check responds with 0 for PASS, 1 for FAIL, and 2 for
MARGIN.
Example:
"INIT;*WAI"
'Starts a new sweep and waits for its end.
"CALC2:LIM3:FAIL?"
'Queries the result of the check for limit line 3 in screen B.
Characteristics:
*RST value: -
SCPI: conform
Mode:
A
CALCulate<1|2>:LIMit<1...8>:NAME
<name of limit line>
This command assigns a name to a limit line. If it does not exist already, a limit line with this name is
created.
The name of the limit line may contain a maximum of 8 characters.
Up to 8 limit lines can be defined at the same time. The command is independent of the measurement
window.
Example:
"CALC:LIM1:NAME 'FM1'"
'Assigns the name 'FM1' to limit line 1.
Characteristics:
*RST value: 'REM1' to 'REM8' for lines 1 to 8
SCPI: device-specific
Mode:
A
Mode:
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...